[Tare timeout]
Timeout for a tare/zeroset command that cannot be executed (e.g. due to mechanical
instability of the scale, incorrect lter setting, resolution too high, standstill condition
too strict).
This parameter is given in seconds. The permitted range is: 0.0…[2.5]…25 s.
At 0.0 s taring is only carried out when the scale is already stable.
[Zeroset range]
Determine a ±range around the zero point determined by the dead load during
calibration; within this range
- the displayed gross weight can be set to zero by pressing the zero-setting key (or
by a corresponding external command), and
- automatic zero tracking is active.
Setting range: 0.00…10000.00 d
For use in legal metrology a value ≤2% of the max must be entered, example: 60 d for
3000 e of Class III.
[Zerotrack]
The zero display is automatically maintained within set limits.
Selection: No/Yes
When [No] is selected, the next three parameters are not shown.
When [Yes] is selected, values for the next three parameters must be entered.
[Zerotrack range]
Range within which automatic zerotrack compensates for deviations.
Setting range: 0.25…10000.00 d
For use in legal metrology a value of 0.25 d must be entered.
[Zerotrack step]
If a weight change exceeds the adjusted value, automatic tracking does not function
any more.
Setting range for automatic tracking increments: 0.25…10 d
For use in legal metrology a value of 0.25 d must be entered.
WP-D
Max 200t d= 0.02t
0000
t
Weighing points @admin
1
2
3
Zerotrack time
Overload
Minimum weight
Range mode
1.0 s
9 d
20 d
single range
Default Save