[W&M]
Setting for operation in legal metrology.
Selection: [None], OIML, NTEP (for USA) or NSC (for Australia), see Chapter 4.7.1.1.1.
[Ambient conditions]
This parameter is used to dene the ambient conditions of the scale.
Selection: very stable condition, Stable condition, Unstable condition, Very unstable
condition
[Detect a defective load cell]
Select and conrm these parameters if the maximum simulation of a faulty load cell is
to be performed automatically. A warning symbol is displayed as long as the load cell
is being simulated.
[Unbalanced check deviation]
The plausibility check is activated when the average deviation is > 0%.
The average deviation of the individual load cells is calculated. Monitoring is
indicated via symbol above the weight unit.
Setting range: 0 to 100%.
[Standstill time]
The parameters [Standstill time] and [Standstill range] are used to dene the
standstill of the scale (stable balance).
The [Standstill time] parameter is entered in seconds. The permitted range is:
0.00…2 s.
The time can be entered from 0.00 to 2.00 seconds, but always makes up at least one
measurement time.
WP-D
Max 200t d= 0.02t
0000
t
Weighing points @admin
1
2
3
Standstill range
Tare timeout
Zeroset range
Zerotrack
Zerotrack range
Zerotrack step
1.00 d
2.5 s
50.00 d
Yes
0.25 d
0.25 d
Default Save
[Standstill range]
As long as the weight uctuations remain within this range, the device is determined
to be stable.
The [Standstill range] parameter is entered in "d". The permitted range is:
0.01…10.00 d.
For use in legal metrology, ≤1 d must be selected.