[Zerotrack time]
Time interval for automatic zerotrack.
Setting range: 0.1…25 s
For use in legal metrology a value of 1 s must be entered.
[Overload]
Weighing range above the maximum capacity (Max) without error message.
Setting range: 0…9999999 d
For use in legal metrology a value of max. 9 d = e must be set.
[Min. weight]
Minimum weight at which a print command can be triggered.
Setting range: 0…9999999 d
For use in legal metrology a value of at least 20 d must be set.
[Range mode]
Selection: single range, multi-range, multi-interval
For range selection for scales and settings, refer to Chapter 4.7.1.1.2 and 4.7.1.1.3
[Default] soft key
Settings are reset to factory settings.
4.7.8.3 Calib
In this menu item, calibration is performed for the digital precision pressure load cell
Pendeo
®
Process.
- [Quick install with search load cells and set dead load], see Chapter 4.7.8.3.1.
- [Search for connected load cells], see Chapter 4.7.8.3.2.
- [View and assign load cells], see Chapter 4.7.8.3.3.
- [Calibrate the scale] (load cells), see Chapter 4.7.8.3.4.
- [New], see Chapter 4.7.8.3.4.1.
- [Modify], see Chapter 4.7.8.3.4.2.
- [Assign load cell name], see Chapter 4.7.8.3.5.
- [Service function for load cells], see Chapter 4.7.8.3.6.
- [Corner correction], see Chapter 4.7.8.3.7.
During calibration, no data is changed in the digital load cells. The calibration data and
parameters are saved in the device. The unique serial numbers of the connected load cells
are monitored.
For the calibration the following order must be followed:
- Search for load cells and set dead load, see Chapter 4.7.8.3.2 or
- Search for load cells only, see Chapter 4.7.8.3.3.
- Assign load cells, see Chapter 4.7.8.3.3.
3. Press the [Save] soft key to save the settings.