Settings are locked for W&M WP A.
Continuing will change the calibration
check number 3004174066!
Press▐█, C ▐█, C ▐█, C and then ██▌OK to clear
W&M and settings locked.
Press ███ EXIT to cancel.
This message appears if the [W&M] parameter has been set to another value besides
[none] for at least one weighing point and [Settings locked] has been activated.
Or: Press the EXIT key to cancel the process and return to the BIOS BOOT menu.
ERROR RECOVERY MENU
Select error recovery option
Restore last software from SD-Card
Listen for FlashIt!
Coldstart in safe mode
3.10.1.3.1 Restore from SD card
Replace a faulty application in the BIOS BOOT menu via restore last software from SD
card.
Accessible via the STOP key [Recover]- [Restore last Software from SD-Card] .
3. Press the C key three times and then press the OK key to reset the [W&M] parameter
and [Settings locked].
The calibration check number changes.
4. If all checks have been carried out, the troubleshooting menu (ERROR RECOVERY
menu) appears.
3 Operation and control Process Controller Maxxis 5 PR 5900
EN-51 Minebea Intec