ERROR RECOVERY MENU
Select error recovery option
Restore last software from SD-Card
Listen for FlashIt!
Coldstart in safe mode
3.10.1.3.2 Load software with FlashIt!
Load a faulty application via software with FlashIt! in the BIOS BOOT menu Replace
(Listen from FlashIt).
Accessible via the STOP key [Recover]- [Listen from FlashIt] .
ERROR RECOVERY MENU
Select error recovery option
Restore last software from SD-Card
Listen for FlashIt!
Coldstart in safe mode
3.10.1.3.3 Cold start in safe mode
Repair a faulty application via cold start in the BIOS BOOT menu.
Accessible via the STOP key - [Recover]- [Cold start in safe mode] .
Conrm [Restore last Software from SD-Card] using the ► key/OK.
The last saved software image (BIOS, rmware, and application) is restored in the
device.
Select [Listen from FlashIt] with the ▼ key and conrm with the ► key/OK.
It is now possible to load software on the device via the network using the
"FlashIt!" program, see also Chapter 6.9.3.
Process Controller Maxxis 5 PR 5900 3 Operation and control