17.4 Eect of ambient conditions
17.4.1 Ambient conditions
Reference temperature 23°C
Ambient temperature for operation -10 to +50°C
Ambient temperature for use in legal met-
rology
-10 to +40°C
Power-on temperature 0 to +50°C
Temperature range
Limits for storage/transport -20 to +70°C
Moisture <95%, non-condensing (acc. to
IEC 60068-2)
Protection class Housing: IP30
Front panel: IP65
Altitude <2000 m
17.4.2 Electromagnetic Compatibility (EMC)
All data in compliance with NAMUR NE 21, EN 45501 and EN 61326.
High frequency electromagnetic elds
(801000 MHz)
EN 61000-4-3 10 V/mHousing
Electrostatic discharge (ESD) EN 61000-4-2 6/8 kV
Fast transients (burst) EN 61000-4-4 1 kV
Peak voltages (surge) 1.2/50 µs EN 61000-4-5 1/2 kV
Signal and control
lines
Conducted disturbances by high fre-
quency coupling (0.1580 MHz)
EN 61000-4-6 10 V
Fast transients (burst) EN 61000-4-4 2 kV
Peak voltages (surge) 1.2/50 µs EN 61000-4-5 1/2 V
Conducted disturbances by high fre-
quency coupling (0.1580 MHz)
EN 61000-4-6 10 V
Voltage dips EN 61000-4-11 0/40%
20/200ms
Mains inputs
Mains failure link EN 61000-4-11 20 ms
17.4.3 RF interference suppression
Interference emission According to EN 61326, limit value class A, for
industrial sectors
17 Specications X3 Process Indicator PR 5410
EN-333 Minebea Intec

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Minebea Intec X3 Process Indicator PR 5410 and is the answer not in the manual?

Minebea Intec X3 Process Indicator PR 5410 Specifications

General IconGeneral
BrandMinebea Intec
ModelX3 Process Indicator PR 5410
CategoryTouch Panel
LanguageEnglish