9.5.2.2 Loading the software on your notebook/PC
9.5.3 Updating with a xed IP address
Requirements:
- Device and notebook/PC are connected to a network/each other.
- The automatic address assignment "DHCP" is deactivated in the device and in the
notebook/PC, see Chapter 7.17.6.
- Notebook/PC is set to a xed IP address.
Note:
On a point-to-point connection the device and notebook/PC must be set to a xed
IP address which has the same address range given by the subnet mask.
Example:
PR 5410: IP address 192.24.22.1
Notebook/PC: IP address 192.24.22.2
Device and notebook/PC have the subnet mask 255.255.255.0.
- The "FlashIt!32" program (in a directory on the enclosed CD-ROM) is installed on the
notebook/PC.
- The "FlashIt!32" program is started.
9.5.3.1 Device presettings in the Setup menu
Now the device is ready to load the software.
1. Click the relevant le in the "Explorer" with the mouse and drag it into the [Select]
window, (or use 'Copy and paste').
2. Under [COM-Port] select "Network" and enter the devices IP addresses.
3. Click [Flash] to start the procedure.
As soon as software loading is completed the device will be re-started.
4. Load next le, as described.
9 Extended functions X3 Process Indicator PR 5410
EN-285 Minebea Intec

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Minebea Intec X3 Process Indicator PR 5410 and is the answer not in the manual?

Minebea Intec X3 Process Indicator PR 5410 Specifications

General IconGeneral
BrandMinebea Intec
ModelX3 Process Indicator PR 5410
CategoryTouch Panel
LanguageEnglish