If you press [Undo], changes are not saved and the display returns to the selection menu
for the weighing points.
Leaving the menu is displayed by "Exit calibration".
7.14.15 Parameter Input
The menu is accessible via -[Weighing point]- [Calib]- [Param].
[Measuretime]
Measuring time: The duration of a measurement can be selected.
Selection: 5 ms, 10 ms, 20 ms, 40 ms, 80 ms, 160 ms, <320 ms>, 640 ms, 960 ms,
1280 ms, 1600 ms.
[Digital lter]
Selection of the digital lter (lter characteristic): <o> (no Filter), Bessel, aperiod.
(aperiodic), Butterw. (Butterworth), Tcheby. (Tschebysche)
The following includes examples of interference signals for the dierent lter types:
Bessel lter Aperiodic lter