Changing the dead load (Cd 003)
Saving and exiting
7.5.4 Displaying the calibration data of the internal WP using the front-panel keys
5. Press to change the parameter number ("000" ashes).
6. Press to select the parameter number.
7. Press OK.
dEAdLo appears on the display.
8. Press OK.
LoAd (dead load with empty scale) appears on the display.
9. Press OK.
unLoAd appears on the display.
10. Empty the scale.
11. Press OK.
A weight value for the dead load appears on the display.
12. Press OK.
00000.0 kg appears on the display.
The dead load is set.
13. Press OK.
Cd 004 appears on the display.
14. Press "Exit."
SAVE appears on the display.
15. Press OK.
YES appears on the display.
16. Press OK.
SAVE ... appears on the display during the save process.
17. Press "Exit" to exit the Setup menu.
18. If no other calibration data or parameters are going to be set, the CAL switch must be
closed now; see Chapter 7.1.3.1.
1. Press .
SEtuP appears on the display.
7 Commissioning X3 Process Indicator PR 5410
EN-131 Minebea Intec

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Minebea Intec X3 Process Indicator PR 5410 and is the answer not in the manual?

Minebea Intec X3 Process Indicator PR 5410 Specifications

General IconGeneral
BrandMinebea Intec
ModelX3 Process Indicator PR 5410
CategoryTouch Panel
LanguageEnglish