Message Possible causes
This message displays, if the scale is not stable.
Remedy
- Check the mechanical function of the scale.
- Adapt the lter setting; reduce the resolution.
- Adapt the standstill conditions.
This message is displayed if the weight on the scale is < the dead load
after input of the weight value.
The maximum capacity is not an integer multiple of the scale interval.
16.5 General error messages
Display Error and Possible Cause
Err 21 BIOS is too old for the rmware.
Err 22 Firmware is too old for the BIOS.
Err 23 Boot not possible:
- ROM error
- No rmware found
Err 25 Plug-in card combination is not permissible.
Err 29 "Flash" process has failed.
Err 31 The digit group for IP address/mask is not within 0–255, or is invalid.
Err 32 The IP address is already in use.
Two instruments with identical IP addresses.
Err 33 Wrong parameter
Err 34 No Ethernet cable (LINK) found
Err 41 Major system error
Err 42 Major system error in controls.
Err 43 Monitoring time has expired.
Err 44 Cannot save to EAROM.
X3 Process Indicator PR 5410 16 Error messages