Message Possible causes
This message is displayed if the selected resolution is so high that less
than 0.8 internal counts per scale interval (d) and/or 0.5 µV/e are avai-
lable if legal-for-trade acc. to OIML/NSC.
This message displays if the maximum load is not an integer multiple of
the scale interval.
Weight units don't match, e.g., subsequent change for [Max] from kg in-
to lb.
This message is displayed when the dead load entered in mV/V plus
scale interval in mV/V is higher than 3 mV/V (= 36 mV).
This message displays, if the scale is not stable.
Remedy
- Check the mechanical function of the scale.
- Adapt the lter setting; reduce the resolution.
- Adapt the standstill conditions.
This message is displayed when the measurement signal is negative
when determining the dead load with [by load].
Cause
Load cell connected with wrong polarity, or defective.
This message is displayed when the dead load entered is 5 mV/V.
16 Error messages X3 Process Indicator PR 5410
EN-327 Minebea Intec

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Minebea Intec X3 Process Indicator PR 5410 and is the answer not in the manual?

Minebea Intec X3 Process Indicator PR 5410 Specifications

General IconGeneral
BrandMinebea Intec
ModelX3 Process Indicator PR 5410
CategoryTouch Panel
LanguageEnglish