9 Extended functions
9.1 Hardware test
9.1.1 Display test
9.1.2 Front-panel key test
9.1.3 Serial interfaces
9.1.3.1 RS-232 interface
Open the menu with - [HW-Slots].
1. Press .
2. Press 2×.
dSP.tSt. appears on the display.
3. Press OK.
When no key has been pressed for 5 seconds, the test ends automatically.
The device goes back to the weight display.
Operate all display elements.
1. Press and hold and also press .
A number is displayed for each key. The number increases by 1 each time a key is
pressed.
0.0.0.0.0.0.
2. For example, press .
0.0.0.1.0.0. appears on the display.
3. For example, press and at the same time.
When no key has been pressed for 5 seconds, the test ends automatically.
The device goes back to the weight display.
-8023- appears on the display.