4.5 Connecting analog load cells and platforms
4.5.1 General information
Load cells or analog platforms (e.g. from the CAPP series) can be connected.
The supply voltage is protected against short circuit and overload.
Note:
The colors listed here apply for the Minebea Intec load cell and connection cables of
type "PR "
Color code
bk = Black
bu = Blue
gn = Green
gy = Gray
rd = Red
wh = White
For additional information on the connection of load cells and cable junction boxes,
refer to the corresponding installation manuals.
4.5.2 Connecting a load cell with a 4-wire cable
The following links between the terminal contacts are provided:
from Sense S+ to Supply V+
from Sense S- to Supply V-
4 Device installation X3 Process Indicator PR 5410