Leaving the menu is displayed by "Exit calibration".
After nishing the calibration, set the CAL switch to the closed position; see also Chapter
7.1.3.1.
7.14.14 Cancelling a calibration
Quit calibration by pressing the softkey.
You are prompted to conrm whether calibration should be closed without determining
the test value.
If not all data was determined when calibrating with [New] (e.g.dead load not set/
entered), this message is shown:
Press [Yes] to conrm and then press again the softkey, another prompt is displayed:
X3 Process Indicator PR 5410 7 Commissioning