[Tare timeout]
Timeout for a zeroset or tare command to be executed.
If the xBPI scale has not executed the command in the specied time, the action will
be aborted.
Setting range: 09.9 s
[Serial number]
Serial number of the connected xBPI scale/weighing module.
The number is required for checking when used in legal metrology.
With serial number "0", checking is omitted.
Setting range: 099999999
[SBN Address]
When the address is not set to 0, bus operation is active. Possible addresses: 1–31,
i.e., max. 31 xBPI scales can be operated on an RS-485 branch.
The SBN Address is shown on the display.
Example: Address 31 at WP-A
7.15.4 Setting up an xBPI platform
3. Enter the following parameters.
4. Press to exit the menu and to save the settings.
1. Select -[Weighing point]- [xBPI-Scale] and conrm.
7 Commissioning X3 Process Indicator PR 5410
EN-181 Minebea Intec