7.6.2 Deleting alibi entries
7.7 Finding and connecting a device with a notebook/PC
If the device is connected to a notebook/PC via a point-to-point connection, an IP
address is negotiated via function "AutoIP". This can take up to 2 minutes!
NOTICE
When the IT/DHCP network cable is temporarily connected between the notebook/
PC and a device, the DHCP server is lost and the notebook/PC returns to the auto-IP
address within approx. two minutes!
Reason: The DHCP server/client relationship is checked cyclically at 23-minute
intervals.
no appears on the display.
17. Press to select "YES."
18. Press OK.
The alibi entry is printed out.
19. Press "Exit" to exit the Setup menu.
1. Press .
SEtuP appears on the display.
2. Press OK.
The "Cd" calibration menu ashes.
Cd 000 appears on the display.
3. Press multiple times to select "AL 100."
4. Press to change the parameter number ("100" ashes).
5. Press to select parameter number "101."
6. Press OK.
ErASE appears on the display.
7. Press OK.
no appears on the display.
8. Press to select "YES."
9. Press OK.
All alibi entries are deleted.
10. Press "Exit" to exit the Setup menu.
X3 Process Indicator PR 5410 7 Commissioning