16.2 Error messages for xBPI scales
Display VNC text Error and possible cause Remote dis-
play
Error 3 Ext. meas. device error No weight values can be read from scale:
- Scale error
Error 3
Error 4 Value exceeds display The weight value is not displayed:
- Too many digits have been set.
Error 4
Error 9 No values from scale No communication with xBPI scale:
- Cable break.
- Internal scale error.
- The scale is not connected to the power sup-
ply.
Error 9
Error 11 No weight data No weight display:
- Another weighing point was selected.
Press to assign the new weighing point to
the device.
Requirements:
- [Serial ports parameter]- [xBPI port]- [Slot
1/2 RS 485] is selected.
Error •
Error 12 Scale not ready The scale is not ready for weighing:
- The device is in the warm-up phase.
- The device is in automatic taring mode.
- The device was switched on with the scale
loaded.
Switch the device o and on again.
Error <
Error 15 Wrong serial number Serial number of scale does not match the num-
ber set in the device.
Error?
16.3 Error messages for Pendeo load cells
Display VNC text Error and possible cause Remote dis-
play
Error 2 Overload The measuring signal is higher than Max + (x d):
- Wrong setting.
- Too much weight on the scale.
Error 2
Error 4 Value exceeds display The weight value is not displayed:
- Too many digits have been set.
Error 4
16 Error messages X3 Process Indicator PR 5410
EN-325 Minebea Intec