Output if <0: set 0 mA = 0 mA, set <4 mA> =
4 mA, set 20 mA = 20 mA, linear = goes below
4 mA down to the limit (with 420 mA)
Output if >Max Output if >Max: set 0 mA = 0 mA, set 4 mA =
4 mA, set <20> mA = 20 mA, linear = exceeds
20 mA up to the limit
Weight at 0/4 mA Weight value for 0/4 mA output
Weight at 20 mA Weight value for 20 mA output
7.14 Calibrating internal weighing point
7.14.1 General information
For legal-for-trade application select - [Weighing point]- [Calib]- [Param], where the
[W & M] must be set to "OIML" before calibration is started; see Chapter 7.14.15.
The calibration data are protected by the CAL switch (see Chapter 7.1.3.1), which is sealed
in the closed (write-protected) position for legal-for-trade applications.
7.14.2 Displaying calibration data
7.14.2.1 Overwrite protection via a CAL switch
When the CAL switch is closed, a tool tip is displayed.
X3 Process Indicator PR 5410 7 Commissioning
Minebea Intec EN-158