The Data under [Calib] and [Param] is displayed only.
The calibration data and parameters are displayed in the format entered/determined
during calibration.
Note:
[Calibrated at]: CAL weight and corresponding mV/V
After input with mV/V, the full scale interval and the mV/V value entered are displayed.
7.14.2.2 Increased resolution (10-fold)
In the -[Weighing point]- [Calib] menu the weight is displayed with 10-fold resolution
(also with the CAL switch closed) with .
After 5 seconds the display returns to normal resolution. Press if you want to switch
to normal resolution immediately.
If the parameter "OIML" for the weighing point under [W&M] is selected, the weight value
is marked as an invalid weight with the symbol.