7.5.6 Search for Pendeo load cells and set dead load using the front-panel keys
In this menu, a search for connected load cells is conducted and the dead load is set.
The load cell type (Pendeo Truck or Pendeo Process) is automatically detected.
Note:
Assigning the Pendeo load cells is only possible with a notebook/PC; see Chapter 7.16.6.
In the process, the dead load is deleted and a recalibration must be carried out.
Procedure
The "Cd" calibration menu ashes.
Cd 000 appears on the display.
3. Press to change the parameter number ("000" ashes).
4. Press multiple times to select parameter number "003."
5. Press OK.
dEAdLo appears on the display.
6. Press OK.
LoAd appears on the display.
7. Press to select "MvoLt."
8. Press OK.
The mV/V value for the dead load appears on the display.
9. Note the displayed value.
10. Press OK.
Cd 004 appears on the display.
11. Press OK.
SPAn appears on the display.
12. Press OK.
MvoLt appears on the display.
13. Press OK.
The mV/V value for the maximum load appears on the display.
14. Note the displayed value.
15. Press "Exit" to exit the Setup menu.
1. Press .
SEtuP appears on the display.
X3 Process Indicator PR 5410 7 Commissioning
Minebea Intec EN-134