Chapter Six: Operation in Tuning Mode How To Ad
j
ust Base
165
2. Select the
BASE
parameter by scrolling through the parameters with the arrow keys.
The system responds by displaying each of the Tuning parameters as they are scrolled
through.
3. Press the [
ENTER
] key.
The system responds by causing the center display to blink.
4. Enter the Base value. The range is from 0 to 103% of full open, and the default is 0.
A. If the center display shows the correct base value, press the [
ENTER
] key.
B. If the center display is not correct, use the numeral keys and the arrow keys (for
tenths of a percent) to enter the correct Base value, then press the [
ENTER
] key.
The system responds by accepting the Base value.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com