EasyManua.ls Logo

MKS 146C - Table 50: Default Values for Dual Channel Displays

MKS 146C
358 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter Seven: Operation in Setup Mode How To Set U
p
a Dual Channel Dis
p
la
y
205
Note
There are no default values supplied for a hot cathode gauge. If 0 is
entered for a switchpoint, the system accepts the value of 0. The default
values for a cold cathode gauge are good values to reference when
determining the best switchpoint values for a hot cathode gauge.
Default Values For Dual Channel Displays
The default value is enabled when 0 is entered
except
where indicated by an
X .
The
X
indicates where the system accepts a value of 0 for the switchpoint.
The upper number is the high switchpoint
The lower number is the low switchpoint
Hi Dual Channel Sensor
(min. resolution for CM)
Lo Dual Channel Sensor
(max. range for CM)
Cold
Cathode
Hot
Cathode
0.1 Torr
CM
1 Torr
CM
10 Torr
CM
100 Torr
CM
HPS
Pirani
9 mT
1 mT
X
100 mT
10 mT
1 T
100 mT
10 T
1 T
X
G.P. Convectron
HPS CEP
9 mT
1 mT
X
100 mT
10 mT
1 T
100 mT
10 T
1 T
100 T
10 T
0.001 mT Linear CM 1 mT
0.1 mT
XXX
10 T
1 T
X
0.01 mT Linear CM 1 mT
0.1 mT
X
100 mT
50 mT
XX X
0.1 mT Linear CM 3 mT
0.3 mT
X
100 mT
50 mT
1 T
0.5 T
XX
1 mT Linear CM 9 mT
1 mT
X
100 mT
50 mT
1 T
0.5 T
10 T
5 T
X
10 mT Linear CM
XX
100 mT
50 mT
1 T
0.5 T
10 T
5 T
100 T
50 T
100 mT Linear CM
XX
100 mT
50 mT
1 T
0.5 T
10 T
5 T
100 T
50 T
107 CM 9 mT
1 mT
X
100 mT
50 mT
1 T
0.5 T
10 T
5 T
100 T
50 T
Table 50: Default Values for Dual Channel Displays