EasyManua.ls Logo

MKS 146C - Sensor Configuration Messages; Table 59: Sensor Configuration Messages

MKS 146C
358 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter Nine: RS-232 Communications Command Messa
g
es
263
Sensor Configuration Messages
The sensor configuration messages set the parameters specific to each type of channel. For
example, you must set the gas type for a Pirani or convection gauge. Table 60, page 264, lists
the parameters for each gauge type.
Sensor Configuration Messages
Message Description Message Format
@
command
category
command
number
Parameter Data
Optional
checksum
Channel Range @ 0 6 1
1, 2, 3, or 4
(channel #)
: ASCII value CR
c1 c2
Channel Resolution @ 0 6 2
1, 2, 3, or 4
(channel #)
: ASCII value
that represents
the resolution
as a power of
10
CR
c1 c2
Channel Gas Type @ 0 6 3
1, 2, 3, or 4
(channel #)
: N = Nitrogen
A = Argon
H = Helium
CR
c1 c2
Channel Gauge
Factor/Sensitivity
@0 6 4
1, 2, 3, or 4
(channel #)
: ASCII value CR
c1 c2
Ion Gauge Disconnect
Threshold
@0 6 5
1, 2, 3, or 4
(channel #)
: ASCII value CR
c1 c2
Hot Cathode High
Pressure Shutoff
@ 0 6 6 1, 2, 3, or 4
(channel #)
: ASCII value CR
c1 c2
Hot Cathode Fast
Pressure Rate of Rise
Shutoff
@ 0 6 7 1, 2, 3, or 4
(channel #)
: E = Enabled
D=Disabled
CR
c1 c2
Convection Gauge
Type
@ 0 6 8 1, 2, 3, or 4
(channel #)
:4=GP
Convectron
5=HPS
Convection
CR
c1 c2
Table 59: Sensor Configuration Messages
Table 61, page 264, lists the chapter that explains the function performed by each message.