Chapter Seven: Operation in Setup Mode How To Perform Sensor Calibration/MFC Setu
199
5. Enter a disconnect threshold for the sensor in the center display.
A. If the display is correct, press the [
ENTER
] key.
B. If the display is incorrect, enter the correct disconnect threshold, then press the
[
ENTER
] key.
The system responds by accepting the new disconnect threshold. The factory
configuration has this feature disabled (the disconnect threshold is set to -1.0).
Note
When this feature is enabled, and the sensor
ressure readin
dro
s below
the disconnect threshold value entered here, an alarm sounds. The factor
configuration value for the sensor disconnect threshold is -1.0, to disable
the feature.
Do not enable this feature if
our
ressure si
nal is nois
in its low ran
e.
The noise may cause the pressure reading to drop below the disconnect
ressure momentaril
; the 146 unit will
ower down the sensor and re
ort
a disconnected sensor status for the channel.