Error Code Summary A
pp
endix C: RS-232 Error Codes
306
E122: Invalid Data Field
An invalid data field error message indicates a problem with the data field of the command
string. The data field can contain a maximum of 26 characters. If your data field exceeds 26
characters, the 146 instrument will return an invalid command error message. Other possible
causes of an invalid command error message include an invalid data range, or a data field that
contains numbers instead of letters, or vice versa.