7-12 Error Messages
6-00054-02 Rev A
60 Cannot complete Lock/Unlock I/E station commands.
Action 1
Open and close the I/E Station and observe the I/E locked indicator on the
Operator Panel. The I/E lock LED should remain illuminated while the I/E
Station is locked. See Table 4-1 on page 4-5 for more information on the I/E
locked indicator.
Action 2
If the problem is not corrected or persists, contact ATAC. For contact
information, refer to ADIC Technical Assistance Center on page 1-5.
61 The Insert/Eject Station Door Closed Sensor error threshold is exceeded.
Action 1
Contact ATAC. For contact information, refer to ADIC Technical Assistance
Center on page 1-5.
70 Failures detected in the Gripper Assembly Finger Open operation.
Action 1
Perform a Teach Current operation (Main Menu Service Teach Teach
Current).
Action 2
If the problem is not corrected or persists, contact ATAC. For contact
information, refer to ADIC Technical Assistance Center on page 1-5.
71 Failures detected in the Gripper Assembly Finger Close operation.
Action 1
Reboot the library. Refer to Shutting Down the Scalar 1000 on page 4-9 and
Starting the Scalar 1000 on page 4-7 for more information.
Action 2
Perform a Teach Current operation (Main Menu Service Teach Teach
Current).
Action 3
If the problem is not corrected or persists, contact ATAC. For contact
information, refer to ADIC Technical Assistance Center on page 1-5.
Table 7-1 SAC Reporting (Continued)
Reported
SAC
Perform these actions