Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
1012User Manual 1178.8520.02 ─ 01
Example:
CALC:DELT:FUNC:FIX ON
Switches on the measurement with fixed reference value for all
delta markers.
CALC:DELT:FUNC:FIX:RPO:X 128 MHZ
Sets the frequency reference to 128 MHz.
CALC:DELT:FUNC:FIX:RPO:Y 30 DBM
Sets the reference level to +30 dBm.
Manual operation: See "Defining a Fixed Reference" on page 434
14.8.3.8 Marker Peak Lists
Useful commands for peak lists described elsewhere
●
CALCulate<n>:MARKer<m>:PEXCursion on page 990
●
MMEMory:STORe<n>:PEAK on page 1086
●
Chapter 14.8.3.3, "Configuring and Performing a Marker Search", on page 989
Remote commands exclusive to peak lists
CALCulate<n>:MARKer<m>:FUNCtion:FPEaks:ANNotation:LABel[:STATe]....................... 1012
CALCulate<n>:MARKer<m>:FUNCtion:FPEaks:COUNt?.................................................1013
CALCulate<n>:MARKer<m>:FUNCtion:FPEaks[:IMMediate]............................................ 1013
CALCulate<n>:MARKer<m>:FUNCtion:FPEaks:LIST:SIZE...............................................1013
CALCulate<n>:MARKer<m>:FUNCtion:FPEaks:SORT.................................................... 1014
CALCulate<n>:MARKer<m>:FUNCtion:FPEaks:STATe....................................................1014
CALCulate<n>:MARKer<m>:FUNCtion:FPEaks:X?......................................................... 1015
CALCulate<n>:MARKer<m>:FUNCtion:FPEaks:Y?......................................................... 1015
CALCulate<n>:MARKer<m>:FUNCtion:FPEaks:ANNotation:LABel[:STATe]
<State>
This command turns labels for peaks found during a peak search on and off.
The labels correspond to the marker number in the marker peak list.
Suffix:
<n>
.
Window
<m> Marker
Parameters:
<State> ON | OFF | 0 | 1
*RST: 1
Example:
CALC:MARK:FUNC:FPE:ANN:LAB:STAT OFF
Removes the peak labels from the diagram
Manual operation: See " Display Marker Numbers " on page 462
Analyzing Measurements (Basics)