Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
1151User Manual 1178.8520.02 ─ 01
Example:
INST:COUP:GEN:STAT ON
INST:COUP:GEN:CENT:STAT ON
INST:COUP:GEN:CENT:OFFS 5
Manual operation: See "Generator Frequency Offset" on page 639
INSTrument:COUPle:GENerator:CENTer[:STATe] <State>
Couples the center frequency of the connected signal generator to the R&S FSV/A.
This command requires the INSTrument:COUPle:GENerator:STATe to be ON.
Parameters:
<State> ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
*RST: 0
Example:
INST:COUP:GEN:STAT ON
INST:COUP:GEN:CENT:STAT ON
Manual operation: See "Center Frequency" on page 639
INSTrument:COUPle:GENerator:RLEVel:OFFSet <Level>
Defines a fixed offset to the reference level of the R&S FSV/A for the coupled signal
generator.
This command requires the INSTrument:COUPle:GENerator:STATe and the
INSTrument:COUPle:GENerator:RLEVel[:STATe] to be ON.
Parameters:
<Level> Default unit: DB
Example:
INST:COUP:GEN:STAT ON
INST:COUP:GEN:RLEV:STAT ON
INST:COUP:GEN:RLEV:OFFS 5
Manual operation: See "Generator Reference Level Offset" on page 640
INSTrument:COUPle:GENerator:RLEVel[:STATe] <State>
Couples the reference level of the connected signal generator to the R&S FSV/A.
This command requires the INSTrument:COUPle:GENerator:STATe to be ON.
Parameters:
<State> ON | OFF | 0 | 1
OFF | 0
Switches the function off
Configuring the R&S FSV/A