Measurements and Results
R&S
®
FSVA3000/ R&S
®
FSV3000
249User Manual 1178.8520.02 ─ 01
Child node Attribute Value Parameter description Mand.
Limit
dBm/Hz | dBm |
dBc | dBr | dB
A Range must contain
exactly two limit nodes; one
of the limit nodes has to
have a relative unit (e.g.
dBc), the other one must
have an absolute unit (e.g.
dBm)
Yes
Start Value <numeric_value> Power limit at start fre-
quency
Yes
Unit dBm/Hz | dBm |
dBc | dBr | dB
Sets the unit of the start
value
Stop Value <numeric_value> Power limit at stop fre-
quency
Unit dBm/Hz | dBm |
dBc | dBr | dB
Sets the unit of the stop
value
LimitFailMode
Absolute | Relative
| Absolute and Rel-
ative | Absolute or
Relative
If used, it has to be identical
to DefaultLimitFailMode
No
RBW Bandwidth <bandwidth in Hz> " RBW " on page 223 Yes
Type NORM | PULS |
CFIL | RRC
No
VBW Bandwidth <bandwidth in Hz> " VBW " on page 223 Yes
Detector
NEG | POS |
SAMP | RMS |
AVER | QUAS
If used, it has to be identical
in all ranges.
No
Sweep Mode Manual | Auto " Sweep Time Mode "
on page 223
Yes
Time <time in sec> " Sweep Time "
on page 224
No
Amplitude
No
ReferenceLevel Value <power in dBm> " Ref Level " on page 224 Yes, if the Refer-
enceLevel child
node is used
Unit dBm Defines dBm as unit Yes, if the Refer-
enceLevel node
is used
RFAttenuation Mode Manual | Auto " RF Att Mode "
on page 224
Yes, if the Refer-
enceLevel child
node is used
Preamplifier
ON | OFF | 1 | 0 " Preamp " on page 224 Yes
Spectrum Emission Mask (SEM) Measurement

Table of Contents

Other manuals for R&S FSV3000 Series