Common Measurement Settings
R&S
®
FSVA3000/ R&S
®
FSV3000
337User Manual 1178.8520.02 ─ 01
For more information on configuring interfaces, see Chapter 13.1.1, "Remote Control
Interfaces and Protocols", on page 644.
Generator Type .......................................................................................................... 337
Interface ..................................................................................................................... 337
TTL Handshake ..........................................................................................................337
GPIB Address / TCPIP Address / Computer Name ................................................... 337
Reference ...................................................................................................................338
Edit Generator Setup File ...........................................................................................338
Frequency Min / Frequency Max ................................................................................338
Level Min / Level Max ................................................................................................ 338
Generator Type
Selects the generator type and thus defines the generator setup file to use.
For an overview of supported generators, see "Overview of Supported Generators"
on page 328. For information on generator setup files, see "Generator Setup Files"
on page 329.
Remote command:
SYSTem:COMMunicate:RDEVice:GENerator<gen>:TYPE on page 936
Interface
Type of interface connection used. The following interfaces are currently supported:
GPIB
TCP/IP (not by all generators)
For details on which signal generators support which interfaces, see the documenta-
tion of the corresponding signal generator.
Remote command:
SYSTem:COMMunicate:RDEVice:GENerator<gen>:INTerface on page 935
TTL Handshake
If available for the specified generator type, this option activates TTL synchronization
via handshake for GPIB connections.
Using the TTL interface allows for considerably higher measurement rates than pure
GPIB control, because the frequency stepping of the R&S FSV/A is directly coupled
with the frequency stepping of the generator.
For more information on TTL synchronization, see "TTL synchronization" on page 333.
For an overview of which generators support TTL synchronization see "Overview of
Supported Generators" on page 328.
Remote command:
SYSTem:COMMunicate:RDEVice:GENerator<gen>:LINK on page 935
GPIB Address / TCPIP Address / Computer Name
For LAN connections: TCP/IP address of the signal generator
For GPIB connections: GPIB address of the signal generator.