Common Analysis and Display Functions
R&S
®
FSVA3000/ R&S
®
FSV3000
495User Manual 1178.8520.02 ─ 01
power measurements, are based on the smoothed trace data. Thus, the results will dif-
fer from results based on the original trace.
For more information see "Trace Smoothing" on page 491.
Remote command:
DISPlay[:WINDow<n>][:SUBWindow<w>]:TRACe<t>:SMOothing[:STATe]
on page 962
DISPlay[:WINDow<n>][:SUBWindow<w>]:TRACe<t>:SMOothing:APERture
on page 962
Average Mode
Defines the mode with which the trace is averaged over several sweeps.
This setting is generally applicable if trace mode "Average" is selected. For FFT
sweeps, the setting also affects the VBW (regardless of whether or not the trace is
averaged).
(See also "Video Bandwidth (VBW)" on page 152).
How many sweeps are averaged is defined by the " Sweep/Average Count "
on page 386.
For details see "How Trace Data is Averaged - the Averaging Mode" on page 491.
"Linear"
The power level values are converted into linear units prior to averag-
ing. After the averaging, the data is converted back into its original
unit.
"Logarithmic"
For logarithmic scaling, the values are averaged in dBm. For linear
scaling, the behavior is the same as with linear averaging.
"Power"
Activates linear power averaging.
The power level values are converted into unit Watt prior to averag-
ing. After the averaging, the data is converted back into its original
unit.
Use this mode to average power values in Volts or Amperes correctly.
In particular, for small VBW values (smaller than the RBW), use
power averaging mode for correct power measurements in FFT
sweep mode.
Remote command:
[SENSe:]AVERage<n>:TYPE on page 963
Average Count
Determines the number of averaging or maximum search procedures If the trace
modes "Average" , "Max Hold" or "Min Hold" are set.
In continuous sweep mode, if sweep count = 0 (default), averaging is performed over
10 sweeps. For sweep count =1, no averaging, Max Hold or Min Hold operations are
performed.
Remote command:
[SENSe:]AVERage<n>:COUNt on page 963
Predefined Trace Settings - Quick Config
Commonly required trace settings have been predefined and can be applied very
quickly by selecting the appropriate button.