Common Analysis and Display Functions
R&S
®
FSVA3000/ R&S
®
FSV3000
494User Manual 1178.8520.02 ─ 01
"Min Hold"
The minimum value is determined from several measurements and
displayed. The R&S FSV/A saves each trace point in the trace mem-
ory only if the new value is lower than the previous one.
The "Detector" is automatically set to "Negative Peak" .
This mode is not available for statistics measurements.
"Average"
The average is formed over several sweeps.
The Sweep/Average Count determines the number of averaging pro-
cedures.
The "Detector" is automatically set to "Sample" .
This mode is not available for statistics measurements.
"View"
The current contents of the trace memory are frozen and displayed.
"Blank"
Removes the selected trace from the display.
Remote command:
DISPlay[:WINDow<n>]:TRACe<t>:MODE on page 959
Detector
Defines the trace detector to be used for trace analysis.
For details see "Mapping Samples to sweep Points with the Trace Detector"
on page 486.
"Auto"
Selects the optimum detector for the selected trace and filter mode.
This is the default setting.
"Type"
Defines the selected detector type.
Remote command:
[SENSe:][WINDow<n>:]DETector<t>[:FUNCtion] on page 964
[SENSe:][WINDow<n>:]DETector<t>[:FUNCtion]:AUTO on page 964
Hold
If activated, traces in "Min Hold" , "Max Hold" and "Average" mode are not reset after
specific parameter changes have been made.
Normally, the measurement is started again after parameter changes, before the mea-
surement results are analyzed (e.g. using a marker). In all cases that require a new
measurement after parameter changes, the trace is reset automatically to avoid false
results (e.g. with span changes). For applications that require no reset after parameter
changes, the automatic reset can be switched off.
The default setting is off.
Remote command:
DISPlay[:WINDow<n>][:SUBWindow<w>]:TRACe<t>:MODE:HCONtinuous
on page 960
Smoothing
If enabled, the trace is smoothed by the specified value (between 1 % and 50 %). The
smoothing value is defined as a percentage of the display width. The larger the
smoothing value, the greater the smoothing effect.
Note: Effects of smoothing on post-processing functions. Note that in Spectrum mode,
all functions performed after the sweep, such as limit checks, markers, or channel