Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
860User Manual 1178.8520.02 ─ 01
Suffix:
<n>
.
Window
<m> Marker
Manual operation: See " Adjust Settings " on page 290
14.5.10.3 Performing the Measurement
The following commands are required to perform a harmonic distortion measurement:
INITiate<n>[:IMMediate] on page 747, see Chapter 14.5.1, "Performing Mea-
surements", on page 745
14.5.10.4 Retrieving Results
The following commands retrieve the results of the harmonic distortion measurement.
CALCulate<n>:MARKer<m>:FUNCtion:HARMonics:DISTortion?........................................ 860
CALCulate<n>:MARKer<m>:FUNCtion:HARMonics:LIST...................................................860
CALCulate<n>:MARKer<m>:FUNCtion:HARMonics:DISTortion?
This command queries the total harmonic distortion of the signal.
To get a valid result, you have to perform a complete measurement with synchroniza-
tion to the end of the measurement before reading out the result. This is only possible
for single sweep mode.
See also INITiate<n>:CONTinuous on page 746.
Suffix:
<n>
.
Window
<m> Marker
Query parameters:
<Result> TOTal
Return values:
<DistortionPct>
<DistortionDb>
Usage: Query only
CALCulate<n>:MARKer<m>:FUNCtion:HARMonics:LIST
This command queries the position of the harmonics.
To get a valid result, you have to perform a complete measurement with synchroniza-
tion to the end of the measurement before reading out the result. This is only possible
for single sweep mode.
See also INITiate<n>:CONTinuous on page 746.
Configuring and Performing Measurements