Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
974User Manual 1178.8520.02 ─ 01
Parameters:
<Position> Vertical position of the trace in % of the height of the diagram
area.
100 PCT corresponds to the upper diagram border.
Range: -100 to 200
*RST: 50
Default unit: PCT
Example:
CALC:MATH:POS 100
Moves the trace to the top of the diagram area.
Manual operation: See " Trace Math Position " on page 517
CALCulate<n>:MATH<t>:STATe
<State>
This command turns the trace mathematics on and off.
Suffix:
<n>
.
Window
<t> irrelevant
Parameters:
<State> ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
Example:
CALC:MATH:STAT ON
Turns on trace mathematics.
Manual operation: See " Trace Math Function " on page 517
See " Trace Math Off " on page 517
14.8.2.4 Retrieving Trace Results
This chapter describes how to retrieve data from standard traces.
For spectrograms see also Chapter 14.8.3.6, "Marker Search (Spectrograms)",
on page 1001.
For details on the format of the retrieved trace data see also Chapter 14.8.2.5, "For-
mats for Returned Values: ASCII Format and Binary Format", on page 977.
Commands required to export traces (and other result data) are described in Chap-
ter 14.9.5, "Storing Measurement Results", on page 1085.
FORMat[:DATA]............................................................................................................. 975
TRACe<n>[:DATA]......................................................................................................... 975
TRACe<n>[:DATA]:MEMory?.......................................................................................... 976
TRACe<n>[:DATA]:X?.................................................................................................... 976
Analyzing Measurements (Basics)