Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
986User Manual 1178.8520.02 ─ 01
<m> Marker
Parameters:
<Position> Numeric value that defines the marker position on the x-axis.
The position is relative to the reference marker.
To select an absolute position you have to change the delta
marker mode with CALCulate<n>:DELTamarker<m>:MODE
on page 984.
A query returns the absolute position of the delta marker.
Range: The value range and unit depend on the measure-
ment and scale of the x-axis.
Default unit: HZ
Example:
CALC:DELT:X?
Outputs the absolute x-value of delta marker 1.
Manual operation: See " Marker 1 / Marker 2 / Marker 3 / Marker 4 " on page 298
See " Marker 1 / Marker 2 / Marker 3 " on page 303
See "Marker Position X-value " on page 431
CALCulate<n>:MARKer<m>:AOFF
This command turns off all markers.
Suffix:
<n>
.
Window
<m> Marker
Example:
CALC:MARK:AOFF
Switches off all markers.
Manual operation: See " All Marker Off " on page 432
CALCulate<n>:MARKer<ms>:LINK:TO:MARKer<md> <State>
This command links normal marker <m1> to any active normal marker <m2>.
If you change the horizontal position of marker <m2>, marker <m1> changes its hori-
zontal position to the same value.
Suffix:
<n>
.
Window
<ms> source marker, see Marker
<md> destination marker, see Marker
Parameters:
<State> ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
Analyzing Measurements (Basics)