Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
999User Manual 1178.8520.02 ─ 01
In the spectrogram, the command moves a marker horizontally to the minimum level in
the currently selected frame. The vertical marker position remains the same.
If the marker is not yet active, the command first activates the marker.
Suffix:
<n>
.
Window
<m> Marker
Manual operation: See " Search Minimum " on page 442
CALCulate<n>:DELTamarker<m>:MINimum:RIGHt
This command moves a delta marker to the next higher minimum value.
The search includes only measurement values to the right of the current marker posi-
tion.
In the spectrogram, the command moves a marker horizontally to the minimum level in
the currently selected frame. The vertical marker position remains the same.
Suffix:
<n>
.
Window
<m> Marker
Manual operation: See " Search Next Minimum " on page 442
14.8.3.5 Retrieving Marker Results
The following commands are used to retrieve the results of markers.
You can use the marker values to position the center frequency or reference level
directly using the following commands:
CALCulate<n>:MARKer<m>:FUNCtion:CENTer on page 885
CALCulate<n>:MARKer<m>:FUNCtion:REFerence on page 899
Useful commands for retrieving results described elsewhere:
CALCulate<n>:DELTamarker<m>:X on page 985
CALCulate<n>:MARKer<m>:X on page 987
CALCulate<n>:MARKer<m>:FUNCtion:FPEaks:COUNt? on page 1013
CALCulate<n>:MARKer<m>:FUNCtion:FPEaks:X? on page 1015
CALCulate<n>:MARKer<m>:FUNCtion:FPEaks:Y? on page 1015
CALCulate<n>:MARKer<m>:FUNCtion:NOISe:RESult? on page 1016
CALCulate<n>:DELTamarker<m>:FUNCtion:PNOise:RESult?
on page 1017
CALCulate<n>:DELTamarker<m>:FUNCtion:BPOWer:RESult?
on page 1022
Analyzing Measurements (Basics)