Measuring range of resistive thermometer Ni x100 Standard
Table 12-10 SIMATIC S7 Format: Measuring Ranges Ni 100, 200, 500, 1000 Standard in °C and °F
Ni x00
Standard in
Units Ni x00
Standard in
Units
°C
(1 digit =
0.1°C)
decimal hexa
decimal
°F
(1 digit =
0.1 °F)
decimal hexa
decimal
Area
> 295,0 32767 7FFF
H
> 563,0 32767 7FFF
H
Overflow
295,0 2950 B86
H
563,0 5630 15FE
H
: : : : : :
Overshoot range
250,1 2501 9C5
H
482,1 4821 12D5
H
250,0 2500 9C4
H
482,0 4820 12D4
H
: : : : : : Nominal range
--60,0 --600 FDA8
H
--76,0 --760 FDA8
H
--60,1 --601 FDA7
H
--76,1 --761 FD07
H
: : : : : :
Undershoot range
--105,0 --1050 FBE6
H
--157,0 --1570 F9DE
H
< --105,0 --32768 8000
H
< --157,0 --32768 8000
H
Underflow
Measuring range of resistive thermometer Ni x00 Climatic
Table 12-11 SIMATIC S7 Format: Measuring Ranges Ni 100, 120, 200, 500, 1000 Climatic in °C and °F
Ni x00
Climatic
Units Ni x00
Climatic
Units
in °C
(1 digit =
0.01°C)
decimal hexa
decimal
in °F
(1 digit =
0.01 °F)
decimal hexa
decimal
Area
> 295,00 32767 7FFF
H
> 325,1 1 32767 7FFF
H
Overflow
295,00
:
250,01
29500
:
25001
733C
H
:
61A9
H
327,66
:
280,01
32766
:
28001
7FFE
H
:
6D61
H
Overshoot range
250,00
:
--60,00
25000
:
--6000
61A8
H
:
E890
H
280,00
:
--76,00
28000
:
--7600
6D60
H
:
E250
H
Nominal range
--60,01
:
--105,00
--6001
:
--10500
E88F
H
:
D6FC
H
--76,01
:
--157,00
--7601
:
--15700
E24F
H
:
C2AC
H
Undershoot range
< -- 105,00 --32768 8000
H
< -- 157,00 --32768 8000
H
Underflow