Analog Electronic Modules
12-86
ET 200S Distributed I/O System
EWA-4NEB 780602402-12
Analog Value Formation
Measurement principle integrating (Sigma-
Delta)
Integration and cycle time /
resolution per channel:
Integration time
selectable
Yes
Interference frequency
suppression in Hz
60 50
Integrationtimeinms 16.7 20
Basic conversion time
incl. integration time in
ms
50 60
Additional conversion
time for wire break
diagnostics in ms
55
Additional conversion
time for wire
compensation with
3-wire connection in ms
50 60
Cycletimeinms Number of active
channels per
module x conversion
time
Resolution (incl.
overshoot range)
Pt100; Ni100; Ni120;
Pt200; Ni200; Pt500;
Ni500; Pt1000;
Ni1000; Cu10 /
15 bits + sign
150 ; 300 ;
600 ; 3000 ; /
15 bits
PTC
1)
/1bit
Interference Suppression, Error Limits
Interference voltage sup-
pression for f= n x (f11%),
(f1 = interference frequen-
cy)
Common mode
interference (U
SS
)
Common mode
interference
(peak value of
interference < rated
value of the input range)
min. 90 dB
min. 70 dB
Crosstalk between inputs min. --50 dB
Operational error limits
(across the temperature
range, in relation to input
range)
Resistive encoder 0.1 %
Pt100, Pt200, Pt500,
Pt1000 Standard
1.0 K
Pt100, Pt200, Pt500,
Pt1000 Climatic
0.25 K
Ni100, Ni120, Ni200,
Ni500, Ni 1000
Standard and Climatic
0.4 K
Cu10 1.5 K
Basic error limit for
resistive sensor (operation-
al error limit at 25 _C, re-
lated to the input range)
Resistive encoder 0.05 %
Pt100, Pt200, Pt500,
Pt1000 Standard
0.6 K
Pt100, Pt200, Pt500,
Pt1000 Climatic
0.13 K
Ni100, Ni120, Ni200,
Ni500, Ni1000 Standard
and Climatic
0.2 K
Cu10 1.0 K
Temperature error (in rela-
tion to input range)
0.0009 %/K
Linearity error (relative to
the input range)
0.01 %
Repetition accuracy
(in transient state at 25_C,
relative to input range)
0.05 %