9-10 Sysmex XP-300
Quality Control
Revised October 2012
Entering from numerical keys dialog
(1) Press the [QC] button at the Ready status.
The analysis control file selection screen will appear.
(2) Press the display column of the file number to be used.
The quality control analysis starting screen for the selected file
will appear.
(3) Press the [Settings] button.
The first QC file setting screen will appear.
When the display column for the parameter to be set is pressed,
the numerical keys will be displayed, and the setting value can
be input.
(4) Press the [Lot ID] column.
The alphanumeric input dialog will appear.
(5) Input the lot ID.
Up to 10 digits can be input for the lot ID.
(6) Press the [Expiration] column.
The date input dialog will appear.
(7) Input the expiration date.
Up to 10 digits can be input for the expiration date.
Note:
QC file settings can be made using file display columns 1 - 6.
If settings are made in a display column that has a registered
QC file, the existing file will be overwritten.
Note:
Pressing the [Setup] button on the QC chart screen also
displays the first QC file setting screen. For further
information regarding the displays, see “9.7 Quality control
chart screen”.
Note:
The analyzer does not judge the expiration dates entered in
the QC file.
Only [yyyy/mm/dd] is available for the expiration date
input format.