Sysmex XP-300 9-17
Quality Control
Revised October 2012
X control
(1) Be sure the status display indicates [Ready] for quality control
analysis.
(2) Mix control blood by rotating and inverting 10 times each for 2
minutes.
(3) Remove the cap while taking care not to allow blood to scatter.
(4) Set the control blood container to the sample probe and press
the start switch in that status.
Note:
QC analysis is always performed in the whole blood mode.
When the analysis mode is set in the pre-diluted mode, the
mode changeover sequence is activated to switch to the
whole blood mode.
Caution!
The control blood (EIGHTCHECK-3WP) remains
usable for the specified period of days in the package
insert. If it is used after this point, abnormal count
values may result.