9-22 Sysmex XP-300
Quality Control
Revised February 2013
The Analyzing screen will appear.
After completion of a single analysis, the results are displayed on the
LCD screen.
The analysis results are displayed in the [Data] column and these
results are compared with the control limit and displayed in the
[Judgment] column on the analysis result screen.
Use the [] or [] button to scroll the screen pages.
To print results on the internal printer, press the [IP] button.
Caution!
When analysis starts, the status reads [Aspirating].
When sample aspiration is completed, the status
display [Aspirating] changes to [Running]. When
[Running] is displayed, the control blood container
can be removed safely.
Do not remove the control blood container from the
sample probe while [Aspirating] is displayed.
Otherwise, there is a possibility that proper aspiration
cannot be performed.
Several seconds after the buzzer sounds “beep,
beep” and [Running] appears on the screen, the
rinse cup lowers. Remove the control blood container
by that time.
To remove the container, lower it straight down. Take
care not to bend the sample probe.
Note:
The sample probe is automatically rinsed, so there is no
need to wipe it clean.
Information
The print format is used for quality control analysis
only and cannot be changed.