Sysmex XP-300 9-23
Quality Control
January 2012
* When the [Top] button is pressed before defining the quality
control data, the quality control analysis confirmation message
“Quit QC Analysis?” will be displayed.
Press the [OK] button to discard all analysis results, complete
the quality control program and return to the Main screen.
Press the [Cancel] button to return to the analysis starting
screen and continue the quality control analysis.
Press the [OK] button to accept the analysis results, and output
the results to the internal printer or host computer, depending on
the quality control data output settings. If the results of the
analysis are not to be accepted, press the [NG] button to
perform a new analysis.
When performing re-analysis, remove the control blood, close
the cap, and mix well. Then, remove the cap again, set the
control blood to the sample probe, and press the start switch of
the instrument.
Example: Sample IDs when quality control analysis was
performed with file No. 3 will be shown as the
following:
Analysis results: QC03
From the stored data screen for output, the result data
can be printed out or transferred to the host computer.
For parameters whose values are beyond the control limit, and if the
values are above the upper limit, is displayed in the [Judgment]
column, and if the values are below the lower limit, is displayed.
Then an alarm sounds and the QC error message is displayed.
Press the button and the action message will appear.
Press the [Back] button to turn off the alarm.
Note:
The analysis results are saved to the stored data.
+
Information
For details about actions to resolve the QC (L-J) error,
see “QC(L-J) Error” in “13. Troubleshooting”.