Sysmex XP-300 10-11
Calibration
Revised October 2012
(5) Enter the calibration value.
Enter the value, and press the [Ent.] button.
The entered value will be defined.
After completion of entering the calibration values, press the
[Quit] button.
The manual calibration setting confirmation dialog will be
displayed.
Press the [OK] button to update the calibration values, and
to return to the Main screen.
Press the [Cancel] button to return to the Main screen
without updating the calibration values.
Note:
If nothing is input (space), the calibration value cannot be
defined.
Information
An alarm sounds and the manual calibration error
confirmation dialog is displayed, if the mean
calibration value for each analysis exceeds 105% or is
less than 95%, or if the new calibration value exceeds
120% or is less than 80%.
The calibration ratio can be calculated using the
following formula.
Calibration Ratio =
Current calibration
value
New calibration
value
× 100