11-4 Sysmex XP-300
Instrument Setup
Revised July 2012
Date/Time
The calendar and clock of the instrument can be set here.
[ISBT128] [Enable]/[Disable
]
When [Enable] is selected, the ISBT128 bar
code will be loaded as a sample ID number with
up to 13 characters.
[ID Inc.] [Enable
]/[Disable]
When [Enable] is selected, the sample ID
number is automatically incremented.
Information
It is not possible to convert the stored data between
Dutch SI units and other units for the following 3
parameters: HGB, MCH and MCHC.
Do not use the stored sample data which was stored
before the setting change. In addition, renew settings
for patient limits, and for quality control TARGET and
LIMIT values.
Parameter Setting
Note:
When time changes to summer or winter time, respectively,
the clock must be manually corrected accordingly.
Numeric input exceeding the upper limit is replaced with
the upper limit value in each parameter.
Parameter Setting
[Format] [yyyy/mm/dd]/[mm/dd/yyyy]/[dd/mm/yyyy
]
[Year] Numeric input: 2000 - 2037
[Month] Numeric input: 1 - 12
[Day] Numeric input: 1 - 28/29/30/31
Determine the available numeric input range by
the Year and Month settings.
[Hour] Numeric input: 0 - 23
[Minute] Numeric input: 0 - 59