Sysmex XP-300 11-5
Instrument Setup
Revised May 2012
Patient Limits
The upper and lower mark limits for the patient results can be entered
here. If the analysis result exceeds the Upper Limit (UL), a [+] flag is
added next to that parameter. If the result is below the Lower Limit
(LL), a [-] flag is added. (See the table below for the initial settings.)
Parameter LL (Lower
Limit)
UL (Upper
Limit)
Unit
WBC 3.0 15.0 × 10
3
/µL
RBC 2.50 5.50 × 10
6
/µL
HGB 8.0 17.0 g/dL
HCT 26.0 50.0 %
MCV 86.0 110.0 fL
MCH 26.0 38.0 pg/dL
MCHC 31.0 37.0 g/dL
PLT 50 400 × 10
3
/µL
RDW-SD 37.0 54.0 fL
RDW-CV 11.0 16.0 %
LYM% (W-SCR) 5.0 55.0 %
MXD% (W-MCR) 1.0 20.0 %
NEUT% (W-LCR) 45.0 95.0 %
LYM# (W-SCC) 0.0 0.0 × 10
3
/µL
MXD# (W-MCC) 0.0 0.0 × 10
3
/µL
NEUT# (W-LCC) 0.0 0.0 × 10
3
/µL
MPV 9.0 13.0 fL
PDW 9.0 17.0 fL
P-LCR 13.0 43.0 %
PCT 0.17 0.35 %
If LL or UL that satify the following conditions was entered
in the Patient Limit setting screen, a beep sounds and the
entering is canceled.
• LL is higher than UL.
• UL is lower than LL.