EasyManua.ls Logo

SYSMEX XP Series - Page 192

SYSMEX XP Series
290 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
12-22 Sysmex XP-300
Cleaning and Maintenance
Revised July 2012
(8) Confirm that the fluid in the TD chamber has been drained.
(9) Remove the TD chamber plug.
(10) Apply CELLCLEAN on the brush provided, and dab by lightly
prodding the brush against the TD aperture.
(11) Mount the TD chamber plug.
(12) Close the transducer cover and main unit front cover. Then turn
on the power.
(13) Check that the blank error has not occurred.
WBC
transducer
RBC transducer
Plug
Caution!
A small volume of reagent occasionally remains.
When reagent has leaked, wipe it off immediately
using a damp cloth.
If left as it is, it can cause current leakage or electric
shocks.
Aperture
Brush
Warning!
CELLCLEAN is a strong alkaline cleaning solution. Do
not mix it with any acidic substances. Also, it should
not come in contact with skin or clothing. If the skin or
clothes should come in touch with it, flush it away
using plenty of water.
Otherwise, it can damage the skin or clothes.
Note:
After using the brush, wash it in water thoroughly to
remove the CELLCLEAN before storing it.
Caution!
The TD chamber plug should be mounted securely.
Otherwise, current leakage or electric shocks can occur.

Table of Contents

Related product manuals