Sysmex XP-300 14-5
Technical Information
Revised March 2012
Linearity:
Whole blood mode
The residual or residual rate against the values analyzed in whole blood mode is
shown.
WBC ±0.3 × 10
3
/µL (1.0 - 9.9 × 10
3
/µL)
±3% (10.0 - 99.9 × 10
3
/µL)
RBC ±0.03 × 10
6
/µL (0.3 - 0.99 × 10
6
/µL)
±3% (1.00 - 7.00 × 10
6
/µL)
HGB ±0.2 g/dL (0.1 - 9.9 g/dL)
±2% (10.0 - 25.0 g/dL)
HCT ±1.0HTC% (10.0 - 33.3%)
±3% (33.4 - 60.0%)
PLT ±10 × 10
3
/µL (10 - 199 × 10
3
/µL)
±5% (200 - 999 × 10
3
/µL)
(However, RBC < 7.00 × 10
6
/µL)
Carryover WBC: 3% or less
RBC: 1.5% or less
HGB: 1.5% or less
HCT: 1.5% or less
PLT: 5% or less
Power supply AC120V/230V/240V±10% (50/60Hz)*
* Refer to the instrument’s name plate for the rated voltage.
Power consumption Approx. 200 VA or less
Heat compensation
required
Approx. 785 BTU/h (198 kcal/h)
Dimensions Main Unit: 420 (W) × 355 (D) × 480 (H) mm
Weight Main Unit: Approx. 30.0 kg
Protection type Class of electric shock protection measures: Class I Equipment
Class of electric shock protection levels: Class B Equipment
EMC
(Electro-magnetic
compatibility)
This equipment is in conformity to the following IEC (EN) standard.
IEC 61326-1: 97+A1: 98 (EN 61326: 97+A1) Electrical equipment for
measurement, control and laboratory use -EMC requirements-
- EMI (Class A)
- EMS (Minimum Immunity test requirements)
EMC Conformity
Assessment
Standards
IEC61326-2-6:2005
EMI:ClassA
Safety IEC61010-1: 2001, IEC61010-2-081: 2001+A1, IEC61010-2-101: 2002