14-4 Sysmex XP-300
Technical Information
Revised July 2012
Precision
(repeatability):
Pre-diluted mode
Values are indicated as variation coefficient (with 95% reliability limit) when
peripheral blood or control blood is analyzed more than 10 times continuously.
WBC 6.0% or less (4.0 × 10
3
/µL or more)
RBC 3.0% or less (4.00 × 10
6
/µL or more)
HGB 2.5% or less
HCT 3.0% or less
MCV 3.0% or less
MCH 3.0% or less
MCHC 3.0% or less
PLT 9.0% or less (100 × 10
3
/µL or more)
LYM% 25.0% or less
MXD% 45.0% or less (12 W-MCR% or more)
NEUT% 25.0% or less
LYM# 25.0% or less
MXD# 45.0% or less (1.0 × 10
3
/µL or more)
NEUT# 25.0% or less
RDW-CV 6.0% or less
RDW-SD 6.0% or less
PDW 18.0% or less
MPV 7.5% or less
P-LCR 30.0% or less
PCT 13.5% or less
Accuracy:
Whole blood mode
Mean values when calibrator is analyzed more than 10 times continuously are
within the following ranges against the displayed values.
WBC ±3% or ±0.2 × 10
3
/µL
RBC ±2% or ±0.03 × 10
6
/µL
PLT ±5% or ±10 × 10
3
/µL
Accuracy:
Pre-diluted mode
Mean values when calibrator is analyzed more than 10 times continuously are
within the following ranges against the displayed values.
WBC ±5% or ±0.3 × 10
3
/µL
RBC ±3% or ±0.05 × 10
6
/µL
PLT ±8% or ±15 × 10
3
/µL