Sysmex XP-300 14-3
Technical Information
Revised July 2012
Precision
(repeatability):
Whole blood mode
Values are indicated as variation coefficient (with 95% reliability limit) when
peripheral blood or control blood is analyzed more than 10 times continuously.
WBC 3.5% or less (4.0 × 10
3
/µL or more)
RBC 2.0% or less (4.00 × 10
6
/µL or more)
HGB 1.5% or less
HCT 2.0% or less
MCV 2.0% or less
MCH 2.0% or less
MCHC 2.0% or less
PLT 6.0% or less (100 × 10
3
/µL or more)
LYM% 15.0% or less
MXD% 30.0% or less (12 W-MCR% or more)
NEUT% 15.0% or less
LYM# 15.0% or less
MXD# 30.0% or less (1.0 × 10
3
/µL or more)
NEUT# 15.0% or less
RDW-CV 4.0% or less
RDW-SD 4.0% or less
PDW 12.0% or less
MPV 5.0% or less
P-LCR 20.0% or less
PCT 9.0% or less