14-42 Sysmex XP-300
Technical Information
Revised October 2012
1) 1A Histogram with high frequency for LD
RBC, HCT, MCV, MCH, MCHC, PLT, and particle distribution width data parameters are
flagged with RL. MCV and RDW-SD exceed the upper Patient Mark Limits, and MCHC falls
below the lower Patient Mark Limits.
RBC Histogram Error - RL (1A)
2) 1B High frequency for LD with 20% particle distribution width error
RBC, HCT, MCV, MCH, MCHC, and PLT parameters are flagged with RL. Of the particle
distribution width data, RDW-CV is flagged with RL and its data are output. RDW-SD
parameter is flagged with DW and its data are not output.
3) 1C High LD with 2 or more peaks
RBC, HCT, MCV, MCH, MCHC, and PLT parameters are flagged with RL. RDW-CV is
flagged with MP and its data are output, while RDW-SD is flagged with MP but its data are not
output.
4) 2A Frequency high for UD
RBC, HCT, MCV, MCH, MCHC, PLT, and particle distribution width data parameters are
flagged with RU. In this instance, flags to indicate that their data are out of Patient Mark Limits
are added to MCV, MCHC, and RDW-SD.
RBC Histogram Error - RU (2A)
RBC
250200100 [fL]
RBC RL 3.57 [×10
6
/µL]
HGB 11.5 [g/dL]
HCT RL 41.8 [%]
MCV RL +117.1 [fL]
MCH RL 32.2 [pg]
MCHC RL -27.5 [g/dL]
RDW-SD RL +91.7 [fL]
PLT RL 185 [×10
3
/µL]
RBC
250200100 [fL]
RBC RU 3.57 [×10
6
/µL]
HGB 11.5 [g/dL]
HCT RU 41.8 [%]
MCV RU +117.1 [fL]
MCH RU 32.2 [pg]
MCHC RU -27.5 [g/dL]
RDW-SD RU +91.7 [fL]
PLT RU 185 [×10
3
/µL]